Lab instruments for oil and gas. Made in Germany.

Haake C/DC Heating Circulators

These space-saving heating circulators are used to control temperatures in smaller external systems such as density meters, viscometers, photometers, refractometers or similar devices.

 

Overview: Haake C/DC-Heating Circulators with bathes P5/U and B3
   (Types with Bädern B5, B7 and B12 in the following table.)

Technical specifications
acc. to DIN 12876
C10-P5/UDC10-P5/UC10-B3DC10-B3DC30-B3DC50-B3
Working temperature range
with tap water cooling
with other cooling
°C
°C
°C
25 ... 100
20 ... 100
0 ... 100
25 ... 100
20 ... 100
0 ... 100
32 ... 100
20 ... 100
-30 ... 100
32 ... 100
20 ... 100
-30 ... 100
32 ... 200
20 ... 200
-50 ... 200
32 ... 200
20 ... 200
-50 ... 200
Temperature accuracy± K0.040.020.040.020.010.01
Heat capacity 230V / 115VkW1.5 / 1.02.0 / 1.21.5 / 1.02.0 / 1.22.0 / 1.22.0 / 1.2
Pump:    pressure max.
Flow rate max.
mbar
l /min
300
12.5
300
12.5
300
12.5
300
12.5
300
12.5
300
12.5
Bath opening: W x L x Dcm13 x 17 x 1613 x 17 x 1613 x 10 x 1513 x 10 x 1513 x 10 x 1513 x 10 x 15
Bath volumeL553333
Overall dimensions: W x L x Hcm16 x 33 x 3616 x 33 x 3620 x 30 x 3720 x 30 x 3720 x 30 x 3720 x 30 x 37
Net weightkg5.25.47.17.37.37.4
Total wattage 230V / 115VVA = W1550 / 10502050 / 12501550 / 10502050 / 12502050 / 12502050 / 1250
Order-No.
230V / 50 ... 60Hz
115V / 60Hz
  
n/a
n/a
 
n/a
n/a
 
n/a
n/a
 
n/a
n/a
 
n/a
n/a
 
n/a
n/a
Extraplus-Rating *) +++++++++++ ++
*): More information on the Haake ExtraPlus-Rating see general catalogue, page 35

 

Overview: Haake C/DC-Heating Circulators with bathes B5, B7 andB12
   (Types with bathes P5/U und B3 in the upper table.)

Technical specification
acc. to DIN 12876
DC30-B5DC50-B5DL30-B7DC50-B7DL30-B12DC50-B12
Working temperature range
with tap water cooling
wither other cooling
°C
°C
°C
32 ... 200
20 ... 200
-50 ... 200
32 ... 200
20 ... 200
-50 ... 200
30 ... 200
20 ... 200
-50 ... 200
30 ... 200
20 ... 200
-50 ... 200
28 ... 200
20 ... 200
-50 ... 200
28 ... 200
20 ... 200
-50 ... 200
Temperature accuracy± K0.010.010.010.010.010.01
Heat capacity 230V / 115VkW2.0 / 1.22.0 / 1.22.0 / 1.22.0 / 1.22.0 / 1.22.0 / 1.2
Pump:    pressure max.
Flow rate max.
mbar
l /min
300
12.5
300
12.5
300
12.5
300
12.5
300
12.5
300
12.5
Bath opening : W x L x Dcm14 x 15 x 1514 x 15 x 1513 x 10 x 2013 x 10 x 2022 x 14 x 2022 x 14 x 20
Bath volumeL4.54.5771212
Overall dimensions: W x L x Hcm21 x 36 x 3821 x 36 x 3823 x 36 x 4423 x 36 x 4432 x 38 x 4432 x 38 x 44
Net weightkg8.58.59.59.513.213.2
Total wattage 230V / 115VVA = W2050 / 12502050 / 12502050 / 12502050 / 12502050 / 12502050 / 1250
Order-No.
230V / 50 ... 60Hz
115V / 60Hz
  
n/a
n/a
 
n/a
n/a
 
n/a
n/a
 
n/a
n/a
 
n/a
n/a
 
n/a
n/a
Extraplus-Rating *) +++++ ++++++++++++ ++
*): More information on the Haake ExtraPlus-Rating see general catalogue, page 35