Application and selection
Very low temperatures are best reached with both P2-CT90L and -W, the P1/P2-C75P and the DC50-K75 cryostats. The P2-CT50L and -W both deliver a high cooling capacity.
P1-C75P / P2-C75P | P2-CT50L | P2-CT50W | ||
• Compact cryostat down to -75 °C • Can be placed on lab table • 280 watt cooling capacity at 20 °C | • High capacity down to -50 °C • Air-cooled, floor-based unit • Ideal for direct temperature control • 2500 watt cooling capacity at 20 °C • Large bath volume (24 l) | • Big power down to -50 °C • Water-cooled floor-based unit • 5000 watt cooling capacity at 20 °C • Large bath volume (24l) | ||
P2-CT90L | P2-CT90W | P2-CT80L | ||
• High capacity down to -90 °C • Air-cooled, floor-based unit • Ideal for direct temperature control • 1650 watt cooling capacity at 20 °C | • Big power down to -90 °C • Water-cooled, floor-based unit • Ideal for direct temperature control • 1900 watt cooling capacity at 20 °C • 200 watt cooling capacity at -80 °C | • Compact cryostat down to -80 °C • Excellent price-performance ratio • 800 watt cooling capacity at 20 °C |
Overview: DC50-K75 and P1-C75P, P2-C75P
(P2-CT80L, -CT50L, -CT50W, -CT90L and CT90W please see table below)
(P2-CT80L, -CT50L, -CT50W, -CT90L and CT90W please see table below)
Technical data acc. DIN 58966 | Unit | DC50-K75 | P1-C75P | P2-C75P | |||
Cooling capacity at 20/0 °C at -20/-40 °C at -60/-80 °C | W W W | 280/220 180/130 50/- | 280/220 180/130 50/- | 280/220 180/130 50/- | |||
Working temperature range | °C | -75 ... 100 | -75 ... 100 | -75 ... 100 | |||
Temperature accuracy | ± K | 0.05 | 0.02 | 0.02 | |||
Heater capacity | kW | 1.5 | 1.0 | 1.0 | |||
Pump: pressure max. flow rate | mbar l /min | 300 12.5 | 560 24 | 560 24 | |||
Bath opening: W x L x D | cm | 13 x 10 x 20 | 13 x 10 x 20 | 13 x 10 x 20 | |||
Bath volume | L | 6 | 6 | 6 | |||
Overall dimensions: W x D x H | cm | 38 x 46 x 72 | 40 x 51 x 77 | 40 x 51 x 77 | |||
Net weight | kg | 65 | 68 | 68 | |||
Toral wattage | VA = W | 2450 | 2500 | 2500 | |||
Order-No 230V / 50 Hz 220V / 60 Hz | n/a n/a | n/a n/a | n/a n/a | ||||
Extraplus-Rating *) | ++++ | +++++ | +++ +++ | ||||
*): More information on the Haake ExtraPlus-Rating see general catalogue, page 35 |
Overview: P2-CT80L, -CT50L, -CT50W, -CT90L, -CT90W
(DC50-K75 and P1-C75P, P2-C75P please see table above.)
(DC50-K75 and P1-C75P, P2-C75P please see table above.)
Technical data acc. DIN 58966 | Unit | P2-CT80L | P2-CT50L | P2-CT50W | P2-CT90L | P2-CT90W | |
Cooling capacity at 20/0 °C at -20/-40 °C at -60/-80 °C | W W W | 800/750 700/600 500/50 | 2500/1750 1100/300 - | 5000/3000 1900/800 - | 1650/1500 1300/1150 600/170 | 1900/1700 1500/1300 700/200 | |
Working temperature range | °C | -80 ... 100 | -50 ... 100 | -50 ... 100 | -90 ... 100 | -90 ... 100 | |
Temperature accuracy | ± K | 0.01 | 0.01 | 0.01 | 0.01 | 0.01 | |
Heater capacity | kW | 1.0 | 2.0 | 3.0 | 2.0 | 3.0 | |
Pump: pressure max. flow rate | mbar l /min | 560 24 | 560 24 | 560 24 | 560 24 | 560 24 | |
Bath opening: W x L x D | cm | 22 x 14 x 20 | 22 x 27 x 20 | 22 x 27 x 20 | 22 x 15 x 20 | 22 x 15 x 20 | |
Bath volume | L | 12 | 24 | 24 | 15 | 15 | |
Overall dimensions: W x D x H | cm | 42 x 66 x 102 | 50 x 75 x 109 | 50 x 75 x 109 | 50 x 90 x 109 | 50 x 90 x 109 | |
Net weight | kg | 107 | 125 | 180 | 190 | 185 | |
Total wattage | VA = W | 2500 | 3300 | 5800 | 5300 | 5300 | |
Order-No 230 V / 50 Hz 220 V / 60 Hz 380 V / 3 Ph / 50 Hz 220 V / 3 Ph / 60 Hz | n/a n/a n/a n/a | n/a n/a n/a n/a | n/a n/a n/a n/a | n/a n/a n/a n/a | n/a n/a n/a n/a | ||
Extraplus-Rating *) | +++ +++ | +++ +++ | +++ +++ | +++ +++ | +++ +++ | ||